TRANSPORT INCUBATOR (AS-5004)



DESCRIPTION

Alarms
• Skin temp  low / high  • Probe failure   • Low battery  • Power failure • System Failure  • Timer
In case of high temperature i.e. above 380C, heater automatically cuts off thus ensuring baby’s safety.
Can be used on manual mode as well.

Collapsible Stainless Steel Stand
With 2/4 no. gas springs for ambulance and in hospital transport purposes. Gas spring system allow the stand to collapse when the lever is pulled and trolley is pressed from both the sides.

Temperature set range

32 °C-  39 °C

Temp variability

1.0 °C

Temp uniformity

1.0 °C

Air filter

removes airborne particles & dust

Easy  Access
Quick and easy access through the side access door and front port holes. The side door folds down and the matteress retracts out  from the hood to access for  emergency procedures. The door and tubing ports  offer ventilator tubing support .Infant restraint strap attached to the baby bed.

Power Source
Transport incubator  operates on AC or DC power using AC when available or switching to its internal battery when necessary
Battery : The batteries used in the system are 12 volt 18ah. Sealed maintenance free VRLA batteries.
Charge Time : 8-10 hrs. per battery from full discharge.
Hood options : Standard single wall / double wall hood (on demand).
Chassis Body : Plastic (fibre) to shock proof the system

Standard Features
• Single wall Hood • Skin temperature probe / Air temp. Probe  • 2 access doors  •  1 baby belt • Collapsible trolley

Optional
• Examination lamp • 2 disposable infant restraint straps • 1 iris port  • 6 tubing ports • Accessory shelf • Suction, Oxygen cylinder, pressure regulatory assembly, Flow Meter, Remote Controlled movement of the Trolley.

Control Displays

Battery condition status

single led indication of low battery condition on the side panel

Power mode 

illuminates AC, DC or external DC

Air/Baby temp °C

displays incubator air temp / baby temperature


 Dimensions in inches (Approx.)

Height without hood

39″

In collapsible stand mode  

29″

height with hood

49″

Width

23″

Length

53″

Baby bed

20×11″

Weight

75kg

Backup

60-90mns

Size of Castors

5″ PU

Sealed maintenance free batter

VRLA Batteries Charge time : 10hrs.(Approx.)